Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-412 3700AS-901-01 Rev. D/June 2018
trigger.timer[N].clear()
This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Usage
trigger.timer[N].clear()
Trigger timer number (1 to 4)
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
Also see
trigger.timer[N].count (on page 11-412)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event.
Instrument reset
Recall setup
Trigger timer N reset
Create configuration script
Save setup
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
Number of times to repeat the trigger (0 to 1,048,575)
Trigger timer number (1 to 4)
Details
If the count is set to a number greater than 1, the timer automatically starts the next trigger timer
delay at the expiration of the previous delay.
Set the count to zero (0) to cause the timer to generate trigger events indefinitely.
If you use the trigger timer with a trigger model, make sure the count value is the same or more than
any count values expected in the trigger model.
Example 1
print(trigger.timer[1].count)
Read trigger count for timer number 1.