Details
Use this attribute to indicate how many measurements to make on a step when measurements are
needed. This sets the measurement count in the trigger model. During a scan, the Series 3700A
iterates through the sequence event detector and measurement action of the trigger model count
times. After performing count iterations, the Series 3700A returns to check the scan count.
This must be set before the scan is started. Once set, it applies to all scan steps in the list, including
scan steps that exist in the list and any that are added before the scan is started.
All steps take the same number of measurements. When taking multiple measurements, the
measurements may be taken as quickly as possible based on the configuration
(scan.trigger.measure.stimulus = 0) or they may be paced by a trigger
(scan.trigger.measure.stimulus is nonzero).
Example