Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-416 3700AS-901-01 Rev. D/June 2018
print(trigger.timer[1].overrun)
If an event was ignored, the output is true.
If the event was not ignored, the output is
false.
Also see
trigger.timer[N].reset() (on page 11-416)
trigger.timer[N].passthrough
This attribute enables or disables the timer trigger pass-through mode.
Instrument reset
Recall setup
Trigger timer N reset
Create
configuration script
Save setup
Usage
passthrough = trigger.timer[N].passthrough
trigger.timer[N].passthrough = passthrough
The state of pass-through mode; set to one of the following values:
â–ª true: Enabled
â–ª false: Disabled
Trigger timer number (1 to 4)
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
trigger.timer[N].reset() (on page 11-416)
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Usage
trigger.timer[N].reset()
Trigger timer number (1 to 4)