Series 3700A System Switch/Multimeter Reference Manual Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018 11-175
dmm.filter.enable
This attribute enables or disables the averaging filter for measurements of the selected function.
Reset
DMM reset
Recall setup
Create configuration script
Save setup
Usage
value = dmm.filter.enable
dmm.filter.enable = value
Filter measurements setting:
â–ª Enable filter measurements: dmm.ON or 1
â–ª Disable filter measurements: dmm.OFF or 0
Details
The filter enable setting only applies when dmm.func is set to one of the following:
• "accurrent"
• "acvolts"
• "commonsideohms"
• "dccurrent"
• "dcvolts"
• "fourwireohms"
• "temperature"
• "twowireohms"
Querying the setting for any other function will return nil and an error message.
The filter enable setting is saved with the dmm.func function setting, so if you use another function,
then return to the previous function, the filter enable setting you set previously is retained.
Example
dmm.func = "twowireohms"
dmm.filter.type = dmm.FILTER_MOVING_AVG
dmm.filter.count = 3
dmm.filter.enable = dmm.ON
Enable filtered measurements for 2-wire
ohms using a moving average filter type with
a count of 3 for each measurement.
Also see
dmm.configure.recall() (on page 11-165)
dmm.configure.set() (on page 11-166)
dmm.filter.count (on page 11-174)
dmm.filter.type (on page 11-176)
dmm.filter.window (on page 11-177)
dmm.func (on page 11-179)
dmm.reset() (on page 11-219)