dmm.configure.recall()
This function recalls a user or factory DMM configuration and replaces attributes in the present configuration with
attributes from the recalled version.
Details
This command recalls the DMM configuration for one function.
When a configuration is recalled, the function associated with the configuration becomes active.
When you recall a DMM configuration, the existing DMM configuration settings for the function are
replaced by the settings in the recalled configuration. Settings for other functions are not affected. For
example, if the function associated with the configuration was temperature, only temperature settings
are recalled. If a factory configuration is recalled, the function's attributes are set to their factory
default values.
The DMM configuration can be user-defined or factory-defined.
User-defined DMM configurations are set with dmm.configure.set(). The factory-defined DMM
configurations are: