WaveSurfer DSO
WS-OM-E Rev B 79
To Recall Default Settings
1. In the menu bar, touch File; then touch Recall Setup... in the drop-down menu.
2. Touch the button under Recall Default Setup
.
The default settings are as follows:
Vertical Timebase Trigger
50 mV/div 50.0 ns/div DC50 or AC1M (model dependent), C1, 0 mV trigger level
0 V offset 5.0 or 10.0 GS/s
(model dependent)
edge trigger
positive edge
0 s delay Auto trigger mode
Saving Screen Images
You can send images to a hard copy printer or to storage media. Both types of output are done
from the same dialog.
1. In the menu bar, touch Utilities, then Utilities Setup... in the drop-down menu.
2. Touch the Hardcopy tab.
3. Touch the File button
.
4. Touch inside the File Format field and select a graphic file type.
5. Under Colors, touch the Use Print Colors checkbox if you want your waveforms to
print in color with a white background. A white background saves printer toner.
6. Touch inside the Directory field and type in the path to the directory where you want
the image stored, using the pop-up keyboard. Or you can touch the browse button
and navigate there.
7. Touch inside the File Name field and type in a name for your image, using the pop-
up keyboard.
8. Under Include On Print, touch inside the Hardcopy Area field and choose Grid
Area Only to exclude the dialog area from the image DSO Window to include the
dialog area Full Screen to reduce the scope window and print the desktop also
9. Touch the Print Now button
.