6 - 18 Operator’s Manual
6 Image Acquisition
6.9.1 Basic Procedures for iScape Imaging
Perform the following procedure:
1. Connect an appropriate iScape compatible probe. Make sure there is enough coupling gel
along the scan path.
2. Press the user-defined key for “iScape View” or tap [iScape View] on the touch screen (it is
available after enter Power/Color mode).
3. Optimize the 2D mode image:
In the capture preparation status, tap [B] ([Power]/[Color]) page tab to go for the B mode
image optimization. Do measurement or add comment/bodymark to the image if needed.
4. Tap [iScape View] page tab to enter the iScape acquisition preparation status.
5. Tap [Start Capture] or press <Update> on the control panel to begin the capture.
The system enters into image review status when the acquisition is completed. You can
perform operations such as parameter adjusting.
6. Scan slowly to obtain a single extended field of view image. You can also erase and retrace if
the image is not satisfactory.
During image acquisition, none of the parameters are adjustable, and functions such as
measurement, comments and body marks are not available.
A green box on the image indicating the boundary between the merged images and the
unfinished images.
During image slicing, the system gives feedback on the probe’s moving speed in the form
of colors and words. The meanings are as follows:
7. Do one of following to end image capture:
Tap [Stop Capture] on the touch screen.
Press <Update>.
Wait until the acquisition completes automatically.
After the acquisition is completed, the panoramic image will be displayed and the system
enters iScape viewing mode.
6.9.2 Image Review
After the acquisition is completed, the panoramic image will be displayed and the system enters
iScape viewing mode.
6.9.3 Evaluate image quality
Many variables may affect the overall image quality. It is important to evaluate the image content
and quality before an image is used for diagnosis or measurements.
Status ROI Color Tip
Speed too low Blue Moving speed of the probe is too low.
Appropriate Green /
Speed too high Red Moving speed of the probe is too high.

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals