Operator’s Manual 13 - 1
13 Measurement, Comments and
Body Mark
13.1 Measurement
There are general measurements and application measurement. You can perform measurements on
a zoomed image, cine reviewing image, real-time image, or frozen image. For measurement details,
please refer to the Advanced Volume.
WARNING
• Be sure to measure areas of interest from the most optimal image plane to
avoid misdiagnosis from inaccurate measurement values.
• To obtain accurate Doppler flow measurement values, make sure the
transmitting beam is not perpendicular to the flow, otherwise false readings
and potential misdiagnosis may result.
CAUTION
• If an image is unfrozen or the mode is changed during a measurement, the
calipers and measurement data will be cleared from the screen, but the
measurement data will be stored in the report.
• If the system is turned off or [End] is selected during a measurement, the
data not saved will be lost.
• In Dual-B imaging mode, the measurement results of the merged image
may be inaccurate. Therefore, the results are provided for reference only,
not for confirming a diagnosis.
Measurement Accuracy
Table 13-1 Error of 2D Images
Parameter Value Range Error
Distance Full screen Within ±3%
Area Full screen Within ±7%
Circ Full screen Within ±10%
Angle Full screen Within ±3%
Volume Full screen Within ±10%

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals