16 Probes and Biopsy
Operator’s Manual 16 - 13
– Wiping: Use a market disinfection wipe product or sterile disposable lint-free soft cloth
wetted with disinfection spray and wipe all surfaces of the probe for a duration according
to the manufacturer instructions.
4. Rinse the probe thoroughly by using a large amount of clean water (about 7.5 L/2 gallons) at
room temperature for about 30 s to remove the residual disinfectant. Repeat the operation
twice. Or follow the disinfectant manufacturer’s instructions regarding rinsing.
5. Dry the probe by wiping with a piece of clean disposable lint-free soft cloth. Do not dry the
probe by heating.
6. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If
such defects exist, the probe has reached the end of its service life. In this case, stop using it
and contact the Mindray service department.
7. Store the probe in a cool, clean and dry environment. And repeat the cleaning and disinfection
process before the next use.
Compatible Cleaners and Disinfectants
For the detailed information about cleaners and disinfectants, see Mindray Transducer Disinfectant
Recommendation.
16.1.6 Cleaning the probe cable and connector
NOTE:
Do not use cloth with water to clean the probe connector.
Perform the following procedure:
1. Wipe out the dust attached to surface of probe connector and cable.
2. Use soft brush to brush the dust inside probe connector gently.
3. Remained stain or dust attached to surface of cable or surface of connector should be washed
out by cloth with a little soapy water, and then air-dry.
16.1.7 Probe Environmental Conditions
Table 16-2 Probe Environmental Conditions
Probe
Model
Conditions Ambient
temperature
Relative humidity
(no condensation)
Atmospheric
pressure
C5-1s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
SC5-1Ns Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
V11-3s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals