USB serial bus trigger settings panel
NOTE. Requires option SRUSB2.
Field or control Description
Source Select the USB bus on which to trigger.
Trigger On Select the type of information on which to trigger.
Handshake Type Sets the handshake packet type on which to trigger.
Available when Trigger On = Handshake Packet.
Packet Type Sets the special packet type on which to trigger.
Available when Trigger On = Special Packet.
Error Type Sets the error type on which to trigger.
Available when Trigger On = Error.
Address Sets the token packet address pattern on which to trigger. Use in conjunction with the Trigger
When field to specify the exact trigger condition.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Token Packet and Token Type = all except SOF (0101).
Address Low, Address High Sets the boundary address conditions when testing for in-range or out of range conditions.
Available when Trigger On = Token Packet and Trigger When = Inside Range or Outside
Range.
Data Low, Data High Sets the boundary data conditions when testing for in-range or out of range conditions.
Available when Trigger On = Data Packet and Trigger When = Inside Range or Outside
Range.
Token Type Sets the token packet type on which to trigger.
Available when Trigger On = Token Packet.
Endpoint Sets the token packet endpoint pattern on which to trigger.
Use in conjunction with the Trigger When field to specify the exact trigger condition.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Token Packet and Token Type = all except SOF (0101).
Frame Number Sets the frame number pattern on which to trigger. Use in conjunction with the Trigger When
field to specify the exact trigger condition.
Tap the Binary, Hex, or Decimal field and use the A and B knobs to select and change the
values. Or double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Token Packet and Token Type = SOF (0101).
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help 375