T Main is the Main input Mid reference crossing time in the configured direction.
T 2nd is the 2nd input Mid2 reference crossing time in the configured direction.
Time Outside Level measurement algorithm
Time Outside Level is the time the signal remains above the high reference level and/or below the low reference level.
Unit Interval measurement algorithm
Unit Interval is the time difference between two successive bits. This measurement is made on each bit in the record. The bits
are calculated using clock recovery method constant mean.
Jitter measurement algorithms
AC Common Mode
AC Common Mode (Pk-Pk) is the peak-to-peak amplitude of the common mode of the sources. Voltage sources are typically
filtered to include only the frequency components above the cutoff frequency (30 kHz). The filter can be disabled in the
measurement configuration. The measurement can be configured to take one or two sources, and the cutoff frequency may be
enabled or disabled. The application calculates this measurement using the following equation:
Peak-to-Peak(High Pass Filter((Source1 + Source2)/2))
Measurement algorithms
454 MSO54, MSO56, MSO58, MSO58LP, MSO64 Help

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals