2. ENVIRONMENT
Your machine should preferably be installed on a perfectly flat and stable work surface of suitable height (about 70 cm), free from shocks
and vibrations, to benefit from the high accuracy of your edger. Your edger can be placed to the right or left of the tracer. However, leave
enough space between the two devices for the opening of the side covers. The tracer may be installed up to 5 m away from the edger (a 5
m cable is supplied with the edger). Longer cables are also available on an optional basis.
The temperature and the humidity of the room where you use your machine must be within the following ranges of values:
Operation:
temperature between +10 °C and +40 °C
Humidity: between 30% and 75%
Storage:
temperature between -5 °C and 50 °C
Humidity: between 25% and 95%
Altitude: < 2,000 m
Level of pollution: 2
Avoid sudden changes in temperature and humidity and install your system:
in an area not directly exposed to sunlight
away from all heat sources
away from all strong magnetic fields
away from all chemical products, corrosive vapours and liquids
Take care not to block the vents and to leave enough space above the device. Do not place any object on the device. Avoid exposing the
device to vibrations or shocks.
The back of your machine must be at least 20 cm away from a wall.
USER MANUAL> TECHNICAL DATA
185
Pro-E 600_Pro-E 700 > v1 -10.18

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Pro-E 700 edger and is the answer not in the manual?

Essilor Pro-E 700 edger Specifications

General IconGeneral
BrandEssilor
ModelPro-E 700 edger
CategoryLaboratory Equipment
LanguageEnglish