Alarm ranges
The alarm names are listed in the Primary Limits and More Limits
tabs on the Alarm Setup
menu. See the "Super user mode" section
for more information on the alarm default settings.
Alarm Range Increment
Pmax (only high) 12-100 cmH2O
1.2 - 9.8 kPa
12 - 98 mbar, hPa
9-73 mmHg
1 cmH2O
0.1 kPa
1 mbar, 1 hPa
1 mmHg
MV High 0.5 - 30.0, Off l/min 0.5 l/min
MV Low Off, 0.1 - 10.0 l/min 0.1 l/min
TV High 20 -1600, Off ml 20 ml
TV Low Off, 1 - 5 ml
5 - 20 ml
20 - 1500 ml
1 ml
5 ml
20 ml
RR High 2 - 100, Off /min 1 /min
RR Low Off, 1 -99 /min 1 /min
EtCO2 High 0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
0.1%
0.1 kPa
1 mmHg
EtCO2 Low Off, 0.1 - 14.9%
Off, 0.1 - 14.9 kPa
Off, 1 - 114 mmHg
0.1%
0.1 kPa
1 mmHg
FiCO2 High 0.1 - 15%, Off
0.1 - 15, Off kPa
1 - 115, Off mmHg
0.1%
0.1 kPa
1 mmHg
FiO2 High 19 - 100%, Off 1%
FiO2 Low 18 - 99% 1%
EtO2 High 19 - 100%, Off 1%
EtO2 Low Off, 1 - 99% 1%
FiIso High 0.1 - 7.0% 0.1%
FiIso Low Off, 0.1 - 6.9% 0.1%
FiSev High 0.1 - 10.0% 0.1%
FiSev Low Off, 0.1 - 9.9% 0.1%
FiDes High 0.1 - 20.0% 0.1%
FiDes Low Off, 0.1 - 19.9% 0.1%
FiEnf High 0.1 - 7.0% 0.1%
FiEnf Low Off, 0.1 - 6.9% 0.1%
FiHal High 0.1 - 7.0% 0.1%
FiHal Low Off, 0.1 - 6.9% 0.1%
EtIso High 0.1 - 7.0%, Off 0.1%
7 Alarms and troubleshooting
2067226-001 7-13