Aladin cassette installation
1.
Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2. If using Aladin2, unlock the cassette handle before installing it
into the active bay.
3. Insert the cassette into the active bay until a click is heard. An
audible click indicates that the cassette is in position.
4. Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5. The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Store the cassettes in the cassette storage bay when they are not in
use. See the "Vaporizer cassettes" section for additional cassette
information.
Note
Aisys CS²
5-2 2067226-001