Maintenance summary and schedule
These schedules indicate the minimum frequency of maintenance
based on typical usage of 2000 hours per year. Service the
equipment more frequently if it is used more than the typical yearly
usage.
Local policies or regulations may require that maintenance be
performed more frequently than stated here.
.
Minimum Frequency Maintenance
Daily
• Clean the external surfaces.
•
Perform a flow and pressure calibration.
• Empty the water reservoir and replace the absorbent in the canister.
Monthly
• Perform 21% and 100% O2 cell calibrations.
During cleaning and setup
• Inspect the parts for damage. Replace or repair as necessary.
As necessary
• Install new cylinder gaskets on cylinder yokes.
• Empty the water reservoir and replace the absorbent in the canister.
• Empty and clean the overflow trap on the optional suction regulator.
• Replace the circuit O2 cell. (Under typical use the cell meets
specifications for 1 year.)
• Replace the disposable flow sensor (plastic). (Under typical use the
sensor meets specifications for a minimum of 3 months.)
• Replace the autoclavable flow sensors (metal). (Under typical use the
sensor meets specifications for a minimum of 1 year.)
• Replace the receiver filter (active gas scavenging only).
• Calibrate the airway modules every 6 months or when there are
indications of errors in the gas readings. Calibrate airway modules that
get extensive usage every 2 months.
• Inspect and clean the fan filters (display, lower electrical enclosure,
and airway module).
Authorized service personnel
This is the minimum level of maintenance recommended. Local
regulations may contain additional maintenance requirements.
Comply with all local regulations which meet or exceed this minimum
level of maintenance.
.
Minimum Frequency Maintenance
12 months Have an authorized service representative complete the scheduled service
maintenance checks, tests, calibrations, and parts replacements as defined
in the Technical Reference manual.
Note
Aisys CS²
9-4 2067226-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE auisys cs2 and is the answer not in the manual?

GE auisys cs2 Specifications

General IconGeneral
BrandGE
Modelauisys cs2
CategoryMedical Equipment
LanguageEnglish

Related product manuals