Typical performance
.
M- and E-series CARESCAPE*
CO2 Measurement range 0 to 15 vol% (0 to 15 kPa,
0 to 113 mmHg).
Measurement rise time less than 400 ms
typical.
Accuracy +/- (0.2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 0.2 vol% (O2,
N2O, anesthetic agents).
Measurement range 0 to 15 vol% (0 to 15 kPa,
0 to 113 mmHg).
Measurement rise time less than 260 ms
typical.
Accuracy +/- (0.2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 0.2 vol% (O2,
N2O, anesthetic agents).
O2 Measurement range 0 to 100 vol%.
Measurement rise time less than 400 ms
typical.
Accuracy +/- (1 vol% + 2% of reading).
6 hour drift less than 0.2 vol%
Gas cross effects less than 1 vol% anesthetic
agents, less than 2 vol% N2O.
Measurement range 0 to 100 vol%.
Measurement rise time less than 260 ms
typical.
Accuracy +/- (1 vol% + 2% of reading).
6 hour drift less than 0.3 vol%
Gas cross effects less than 1 vol% anesthetic
agents, less than 2 vol% N2O.
N2O Measurement range 0 to 100 vol%.
Measurement rise time less than 450 ms
typical.
Accuracy +/- (2 vol% + 2% of reading).
6 hour drift less than 0.1 vol%
Gas cross effects less than 2 vol% anesthetic
agents.
Measurement range 0 to 100 vol%.
Measurement rise time less than 320 ms
typical.
Accuracy +/- (2 vol% + 2% of reading).
6 hour drift less than 0.3 vol%
Gas cross effects less than 2 vol% anesthetic
agents.
Anesthetic agents Measurement range Hal, Enf, Iso 0 to 6% vol
%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time less than 400 ms
typical.
Accuracy +/- (0.15 vol% + 5% of reading).
6 hour drift for Hal, Enf, Iso, Sev less than 0.1
vol%
6 hour drift for Des less than 0.2 vol%
Gas cross effects less than 0.15 vol% N2O.
Measurement range Hal, Enf, Iso 0 to 6% vol
%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time Des, Enf, Iso, Sev less
than 420 ms typical.
Measurement rise time Hal less than 800 ms
typical.
Accuracy +/- (0.15 vol% + 5% of reading).
6 hour drift for Hal, Enf, Iso, Sev less than 0.1
vol%
6 hour drift for Des less than 0.3 vol%
Gas cross effects less than 0.15 vol% N2O.
*The effects caused by N2O to the measurement of CO2, O2, and anesthetic agents are automatically
compensated for. The effects caused by anesthetic agents to the measurement of CO2 and N2O are automatically
compensated for.
11 Specifications and theory of operation
2067226-001 11-21