2. Select Volume Apnea Setup.
3.
Set Volume Apnea Selection to Enable or Disable.
Set to Enable to allow the user to turn the volume apnea alarm
on or off from the Start Case menu.
Set to Disable to set the volume apnea alarms as always on
during manual ventilation.
4. Set the Volume Apnea for each case default.
5. Select Back to return to the Case Defaults menu.
When the Volume Apnea Selection is set to Disable, the Volume
Apnea Alarm menu selection does not show on the user’s Start
Case menu and the Vol Apnea Alarm selection does not show on
the Alarm Setup menu.
--
Default case type settings
Each case has multiple settings. The default settings for the default
case types are shown in the following table. An * indicates that the
setting is not used for the default ventilation mode in the case type.
Use the empty columns to write in facility changes.
VCV ADULT settings are used as the default if the system does not
have an optional ventilation mode shown.
.
Page 1 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Name ADULT PEDIATRIC LOCAL CUSTOM 1
Patient and Sensor Adult Pedi Adult Adult
Ideal Weight 70 18 70 70
Age 40 5 40 40
Other Gas Air Air Air Air
Circuit Circle Circle Circle Circle
Data Source Vent Vent Vent Vent
O2% 100 100 100 100
Total Flow 6 6 0.2 6
Auto MV Limits Off Off Off Off
Alarm Volume 3 3 1 3
MV/TV Alarms On On On On
Note
Note
12 Super user mode
2067226-001 12-11