Viewing alarm history
Use the Alarm History
tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
1. Select Alarm Setup.
2. Select the Alarm History tab.
The list of alarms shows in the window.
3. Select Close.
Setting alarm volume
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Alarm Volume to the desired value.
The alarm volume range is 1 to 5.
4. Select Close.
Setting apnea delay
Use the Apnea Delay setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Apnea Delay to the desired time.
• The apnea time delay range is 10 to 30 seconds.
• The apnea time settings are in 1 second increments.
4. Select Close.
Silencing leak audio alarms
Use the Leak Audio setting to silence audio alarms of small leaks.
1. Select Alarm Setup.
2. Select the Configure tab.
3 Operation
2067226-001 3-19