The Pressure Hold, Hold Time, and PEEP on Exit
settings can be
preset by the Super User. These settings can be changed by the
user before starting the procedure.
PEEP on Exit shows if it is set to Yes by the Super User.
Using vital capacity
1. Select Procedures.
2. Select Start Vital Capacity.
One pressure breath is given at the set pressure.
The pressure is held for the set time.
PEEP is set to the PEEP on Exit setting.
3. Select Stop Vital Capacity at any time to stop the procedure.
4. Select Close.
If the procedure is stopped before completion, the PEEP on Exit
setting is not used.
Changing vital capacity settings
1. Select Procedures.
2. Select the setting to change and make the change.
• Set Pressure Hold to between 20 and 60 cmH2O.
• Set Hold Time to between 10 and 40 seconds.
• Set PEEP on Exit to Off or between 4 and 30 cmH2O.
3. Select Start Vital Capacity.
4. Select Close.
Cycling
Use the Cycling procedure to deliver pressure breaths through a
series of ventilation steps. The Cycling procedure provides a flexible
way to deliver pressure breaths during ventilation without making
multiple ventilator setting changes. Up to seven preset steps with
multiple breaths are available.
Each procedure defaults steps and ventilation settings which can be
preset by the Super User. The ventilation settings of each step can
be changed by the user before starting a procedure.
There is a limited amount of gas in the bellows. No additional gas
enters the bellows during the cycling procedure. Increase fresh gas
flow to avoid bellows collapse.
Using cycling
1. Select Procedures.
Note
Note
Note
3 Operation
2067226-001 3-29