Ventilation modes factory default settings
The ventilation mode factory default settings for the ADULT case
default show in the following table. An * indicates the setting is not
used for the ventilation mode. See the
"Super user mode" section for
all other case default values.
.
Setting Range Mode
VCV PCV PCV-VG SIMV VCV SIMV PCV SIMV
PCV-VG
PSVPro CPAP
+PSV
TV 20-1500 ml 500 * 500 500 * 500 * *
RR 4-100 bpm 12 12 12 * * * * *
RR 2-60 bpm * * * 12 12 12 12 *
Minimum RR 4-60 /min * * * * * * * 12
I:E 2:1 - 1:8 1:2 1:2 1:2 * * * * *
Tpause Off, 5-60% of
Tinsp
Off * * Off * * * *
PEEP Off, 4-30
cmH2O
Off Off Off Off Off Off Off Off
Pmax 12-100
cmH2O
40 40 40 40 40 40 40 40
Pinsp 5-60 cmH2O * 5 * * 5 * 5 5
Psupport Off, 2-40
cmH2O
* * * 2 2 2 2 2
Tinsp 0.2-5 seconds * * * 1.7 1.7 1.7 1.7 1.7
Rise Rate Auto, 1-10 * Auto * Auto Auto * Auto Auto
Rise Rate 1-10 * * 5 * * 5 * *
PSV Rise Rate Auto, 1-10 * * * * * Auto * *
Trig Window Off, 5-80% of
Texp
* * * 25 25 25 25 *
Flow Trigger 0.2-10 l/min * * * 2.0 2.0 2.0 2.0 2.0
End of Breath 5-75% of peak
inspiratory
flow
* * * 30 30 30 30 30
Backup Time 10-30 seconds * * * * * * 30 *
Exit Backup Off, 1-5
spontaneous
breaths
* * * * * * 2 *
11 Specifications and theory of operation
2067226-001 11-35