1. Remove the cassette from the machine, and place it on a
horizontal surface. Make sure that the valve pins on the back of
the cassette do not come in contact with the machine or any
other object.
2.
Remove the yellow protective cap from the anesthetic agent
bottle. Check that the filling mechanism on the bottle is not
damaged.
3. Remove the filling port cap from the cassette by turning the cap
counterclockwise.
4. Insert the nozzle of the bottle into the filling port.
5. Push the agent bottle firmly into the filling port. Let the agent
flow into the cassette.
6. Watch the liquid level indicator.
• When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
--
7. Put the cap back on the filling port.
8. Put the cap back on the agent bottle.
Filling with Saf-T-Fil bottle
AB.60.017
Figure 13-12 • Filling Aladin cassette with Saf-T-Fil bottle
Do not attempt to fill the cassette with desflurane that is
warmer than 26C/78F. The overfill prevention system
may lock and prevent the filling of the cassette.
Note
CAUTION
Aisys CS²
13-18 2067226-001