Electronically controlled vaporizer and Aladin cassette
Electronically controlled vaporizer for delivery of five agents:
halothane, isoflurane, enflurane, sevoflurane, and desflurane. The
cassettes can be handled safely without excessive leakage of
anesthetic agent to the environment. The electronically controlled
vaporizer is calibrated using 100% O2 at multiple flows.
.
Agent setting ranges
Halothane, enflurane, isoflurane Off, 0.2 to 5% in fresh gas flow, resolution 0.1%
Sevoflurane Off, 0.2 to 8% in fresh gas flow, resolution 0.1%
Desflurane Off, 1 to 18% in fresh gas flow, resolution 0.2%
.
Response time
To 90% of step; measured at fresh gas outlet Less than 7 seconds at fresh gas flow of 2 l/min
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
11000
12000
13000
14000
15000
18-25C
10-35C
AB.60.055
+20%, +1.0% v/v
1
2
+10%, +0.5% v/v
-10%, -0.5% v/v
-20%, -1.0% v/v
1. Accuracy is greater of percent of setting (%) or volumetric percent (%
v/v)
2.
Fresh gas flow ml/min
3. All settings
4. Settings equal to or less than 9.0%
Figure 11-12 • Desflurane accuracy Aladin and Aladin2 Enhanced
Temperature Sensing
Aisys CS²
11-40 2067226-001