Gas specifications for CARESCAPE modules
.
CARESCAPE
Airway humidity 0 to 100% condensing
Sampling delay 3.0 seconds typical with a 3 m sampling line
Total system response time Less than 3.8 seconds with a 3 m sampling line
Warm-up time 1 minute for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
Respiration rate 4 to 100 breaths/min
Diversion flow 120 +/- 20 ml/min
Airway pressure -20 cmH2O to 100 cmH2O
.
CARESCAPE airway modules: Accuracy under different conditions
Ambient temperature: 10 to 40°C
Ambient pressure: 495 to 795 mmHg
Ambient Humidity: 10 to 98% RH, non-condensing
Automatic compensation for ambient pressure
Full module accuracy for Respiration rate of 4 to 70 breaths/min
During stable condition During warm-up 1 to 10 minutes
(anesthetic agents 5-10 minutes)
During warm-up 10-20 minutes
CO2 +/- (0.2 vol% + 2% of reading) +/- (0.4 vol% + 7% of reading) +/- (0.3 vol% + 4% of reading)
O2 +/- (1 vol% + 2% of reading) +/- (3 vol% + 3% of reading) +/- (2 vol% + 2% of reading)
N2O +/- (2 vol% + 2% of reading)
between 0 and 85 vol%
+/- (2 vol% + 8% of reading for
N2O between 85 and 100 vol%
+/- (3 vol% + 5% of reading) +/- (3% vol% + 3% of reading)
Hal, Enf, Iso,
Sev, Des
+/- (0.15 vol% + 5% of reading) +/- (0.3 vol% +10% of reading) +/- (0.2 vol% +10% of reading
Aisys CS²
11-20 2067226-001