The adjusted MAC value shows on several areas of the screen
including in the mini-trend, agent waveform numeric information,
agent digit field, and graphical trends page.
Starting a case using default settings
Start a case using the default settings by case type defined by the
Super User.
Case Defaults contain five case type selections. Each case type has
preset values for Ideal Weight, Age, and Volume Apnea Alarm.
The first four default case types are configured and named by the
Super User. The fifth default case is Last Case.
1.
Set the Bag/Vent switch to Bag.
2. Select Start Case.
The Case Defaults selection shows the first preset case type.
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm
show the default settings that correspond to the case type
shown.
3. Verify or change the Case Defaults selected.
4. Verify the settings are clinically appropriate.
5. Select Start Case Now. Gas flow starts.
Starting a case using custom settings
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm can be
custom set on the Start Case menu before starting a case.
Additional ventilator settings, ventilation mode, alarm settings, and
gas settings can be custom set through the Vent Mode menu and
other ventilation quick keys, Alarm Setup menu, Gas Setup menu.
1. Set the Bag/Vent switch to Bag.
2. Select Start Case.
The Case Defaults selection shows the first preset case type.
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm
show the default settings that correspond to the case type
shown.
3. Change Ideal Weight, Age, or Volume Apnea Alarm settings
on the menu.
The Case Defaults changes from the case name to Preset.
If the CO2 Alarms setting on the menu is changed, the Case
Defaults remains as previously selected.
3 Operation
2067226-001 3-5