Setting up the absorber canister
The absorber canister is available in two versions: Disposable Multi
Absorber and Reusable Multi Absorber. Both are removed and
installed on the breathing system the same way.
Each canister holds 800 grams of loose absorbent. The
manufacturer recommends Medisorb TM absorbent.
Both absorber versions should only be used with mixtures of air,
oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane
and sevoflurane.
1
2
3
4
5
6
AB.74p042
7
AB.74p.043
1. Canister support pin
2. Canister handle
3. Disposable Multi Absorber canister
4. Absorbent
5. Expiratory water reservoir
6. Canister release latch
8 Setup and connections