Using EZchange canister mode
Make sure that the absorber canister has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
1.
Push the absorber canister release to activate the EZchange
canister mode.
The canister swings down to the EZchange position.
‘CO2 Absorber Out of Circuit’ shows in the waveform area on
the anesthesia display.
2. Remove the canister from the holder.
3. Install a canister with fresh absorbent into the holder.
4. Push the canister back up and snap it into absorber position.
The exhaled gas flows through the absorber, removing CO2.
Note
3 Operation
2067226-001 3-37