Power immunity
.
Power immunity
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
Mains power quality should be that of
a typical commercial and/or hospital
environment.
Surge IEC 61000-4-5 +/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
+/- 1 kV differential mode
+/- 2 kV common mode
Mains power quality should be that of
a typical commercial and/or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in
UT) for 5 seconds
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 seconds
Mains power quality should be that of
a typical commercial and/or hospital
environment. If the user of the system
requires continued operation during
power mains interruptions, it is
recommended that the system be
powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz magnetic
field IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Note: UT is the AC mains voltage prior to application of the test level.
Radiated immunity
.
Radiated immunity
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment
guidance recommended
separation distance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
system, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Aisys CS²
11-48 2067226-001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE auisys cs2 and is the answer not in the manual?

GE auisys cs2 Specifications

General IconGeneral
BrandGE
Modelauisys cs2
CategoryMedical Equipment
LanguageEnglish

Related product manuals