Electrical power
Supply voltage 100-120, 220-240, or 120/220-240 Vac +/- 10% at 50
or 60 Hz
Inlet circuit
breakers
100-120 Vac 220-240 Vac 120/220-240 Vac
15 A 8 A 8 A
Outlet circuit
breakers
110-120 Vac Japan 220-240 Vac
(3) 2 A
(1) 3 A
(2) 2 A
(1) 4 A
(3) 1 A
(1) 2 A
System leakage
current limit - do
not exceed
UL and CSA rated systems (U.S.A. and Canada): less
than 300 μamps for the system and all systems
connected to electrical outlets.
IEC rated systems (not U.S.A. and Canada): less than
500 µamps for the system and all systems connected
to electrical outlets.
Note: Products connected to electrical outlets may
increase the leakage current above these limits.
Resistance to
ground
Less than 0.2 Ω
Power cord
Length 5 meters
Voltage rating 100 to 240 Vac
Current capacity 10 A for 220-240 Vac
15 A for 100-120 Vac
10 A for 120/220-240 Vac
Type Three conduction power supply cord (medical
grade where required).
This equipment must be connected to a supply mains
with a protective earth to avoid the risk of electric shock.
Battery information
The system is not a portable unit. A sealed lead acid battery supplies
backup power in the event of a power failure.
• Capacity to operate for 50 minutes under typical operating
conditions and 30 minutes under extreme conditions.
WARNING
Aisys CS²
11-10 2067226-001