Pneumatic problems
.
Symptom Problem Solution
High-pressure leak test fails. Controls are not set correctly. Make sure no gas is flowing, turn off the
auxiliary flow meter, and repeat the test.
Incorrect cylinder connection. Make sure that there is only one cylinder
gasket, the gasket is in good condition, and
the connection is tight.
Low-pressure leak. Anesthesia machine problem. Contact an authorized service
representative.
7 Alarms and troubleshooting
2067226-001 7-19