COMMISSIONING
136 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Average Temperature Measurement (FII-RTD)
7.3.8.1 Introduction
The Field Interface Instrument - Resistance Temperature Detector (FII-
RTD) board is a sensor module for the instrument (gauge) and
calculates average product, Vapour- and ambient temperatures.
For accomplishing this, RTDs or MRTs must be directly connected to
the FII-RTD module
Figure 7-22
The FII-RTD board
There are four possible configurations to calculate the average product,
Vapour, and ambient temperatures using the FII-RTD board.
ď‚§ 1 or 2 RTDs in a 3-wire configuration.
ď‚§ 1 or 2 RTDs in a 4-wire configuration.
ď‚§ 2 through 6 RTDs as a probe called MPT. It should contain 1
wire per RTD + 2 common wires for all RTDs.
ď‚§ 2 through 6 MRTs as a probe. It should contain 1 wire per
MRT + 2 common wires for all MRTs.
NOTE: Only in RTD 3-wire or 4-wire configurations ambient
temperature calculation is possible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals