MAINTENANCE
266 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Replacing the drum bearings
1. Open the drum cover.
2. Take out the drum shaft bushing.
3. Replace the bearings, refer to figure below
4. Install the bushing.
5. Install drum and displacer.
6. Close the cover.
Figure 9-8 Replacing the drum shaft bearings
Do not grease the carbon PTFE bearings. The drum bearings do not require any
lubricant at all.