MAINTENANCE
Part No.: 4417340 Revision 1 Service Manual 275
954 SmartServo FlexLine
9.6 Synchronizing the reference encoder
After the installation of a new motor block, the internal reference
encoder must be synchronized to the position of the reference encoder
and the gauge starts with an error with status code e.g. ‘Find positions
failed offset to big’ (SmartView code 6158). The following procedure will
solve this issue.
Entity name Entity name
(SmartView)
Description
[Set to
maintenance
mode]
[Maintenance mode]
(Product level)
Go into maintenance mode and do not enter any
command which is not specified below.
[Find reference
encoder position]
[Find ref encoder]
(Product level)
The 954 SmartServo FlexLine finds its encoder
position. Wait approximately 20 seconds.
[Set to
operational
mode]
[Operational mode]
(Product level)
Restart in operational mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals