COMMISSIONING
78 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
7.3.4.10 Special commands
Name Value Range Default Value Explanation
[Interface 1]
This command activates
the Servo to search for
interface 1, the so called
vapor-product interface,
typically called the
product level. It uses
[Setpoint 1] and
[Integration time 1]
settings.
[Interface 2]
This command activates
the Servo to search for
interface 2, the so called
product with density A-
product with density B
interface, typically called
the interface 2 level. It
uses [Setpoint 2] and
[Integration time 2]
settings.
[Interface 3]
This command activates
the Servo to search for
interface 3, the so called
product-water interface,
typically called the water
level. It uses [Setpoint 3]
and [Integration time 3]
settings.
[Dip mode]
This command will set the
displacer at some
distance [Dip Height]
above the interface 1.
After some time [Dip time]
the interface 1 is
searched/dipped and the
displacer will return back
to the dip height position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell FlexLine SmartServo 954 and is the answer not in the manual?

Honeywell FlexLine SmartServo 954 Specifications

General IconGeneral
BrandHoneywell
ModelFlexLine SmartServo 954
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals