COMMISSIONING
90 Service Manual Part No.: 4417340 Revision 1
954 SmartServo FlexLine
Name Value Range Default Value Explanation
[Measured
frequency status]
<active, not
active>
This entity gives the status
whether the F0..F3 are being
measured.
[Measure frequency
F0]
This command will measure the
frequency with the actual
attached weight of 25 g. The
result is placed in entity
[frequency 0].
[Measure frequency
F1]
This command will measure the
frequency with the actual
attached weight of 100 g. The
result is placed in entity
[frequency 1].
[Measure frequency
F2]
This command will measure the
frequency with the actual
attached weight of 175 g. The
result is placed in entity
[frequency 2].
[Measure frequency
F3]
This command will measure the
frequency with the actual
attached weight of 250 g. The
result is placed in entity
[frequency 3].
[Measured
frequency]
This entity holds the measured
frequency after the [Measure
frequency] command has been
given.
[Measured weight] This entity holds the measured
weight of the displacer after the
[Measure frequency] command
has been given.
[Measured
compensated
weight]
This entity holds the measured
weight of the displacer minus
the wire weight after the
[Measure frequency] command
has been given.
[Measured
frequency
continuously]
Shows the continuous measured
frequency
[Go up] This command will let the
displacer run up continuously. It
can only be activated during
maintenance mode.
The command can be
interrupted by Freeze or Go
down command.
No wire tension check!