EasyManuals Logo
Home>Hydac>Accessories>CS1000 series

Hydac CS1000 series User Manual

Hydac CS1000 series
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
Status Messages
Exceptions Errors
LED
Flash code /
display
CS1000 Status
To do
Error
no.
Analog output
Switching output
Flow status on the
digital output
OFF
CS
no display
no function.
Check the supply
voltage for the CSr.
Contact HYDAC.
-
0 mA / 0 V*
open
-
Red
-
Firmware error
Reset the unit. (To
do this, disconnect
the CS from the
voltage supply) or
contact HYDAC.
-1…-19
4.1 mA / 2.05 V*
open
-
Red
-
Communication
error
Check the wiring.
-20…-39
4.1 mA / 2.05 V*
open
-
ContaminationSensor CS1000
en(us)
Page 107 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx 2017-09-25