EasyManuals Logo
Home>Hydac>Accessories>CS1000 series

Hydac CS1000 series User Manual

Hydac CS1000 series
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
CS1000 Characteristics
CS1000 Characteristics
The ContaminationSensor Series CS1000 is a stationary measurement unit
for the continuous monitoring of particulate contamination in hydraulic and
lubrication systems.
The CS is designed to be used in low- or high-pressure hydraulic and
lubrication circuits and test benches where a small amount of oil (between 30
ml/min and 500 ml/min) is diverted for measurement purposes.
The ContaminationSensor is approved for a maximum operating pressure
(see specification on type label) and viscosity of up to 1000 mm²/s.
Particulate contamination is detected with an optical measurement cell
The sensor is available with the following options:
with or without 6-digit display and keypad (can be rotated by 270°)
with 4 - 20 mA or 2 - 10 V analogue output
Results are output as a cleanliness code according to:
4406:1999 and ISOSAE or
AS 4059 4406:1987 and ISO or ISO4406:1999 and NAS SAEAS 4059
pipe / hose installation or flange installation
All models feature an analog electric output and an RS485 interface for
outputting the measured cleanliness class. In addition, all CS1000's have a
switching output.
ContaminationSensor CS1000
en(us)
Page 18 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx 2017-09-25