Setting the measuring mode
Mode M4: Filter to continuously monitor cleanliness class
Control of stationary offline filtration unit
Display & RS485 & analog output & switching output
Establish continuous monitoring of cleanliness class
between min./max. limit values.
Control of a filter unit, continuous measurement of solid
particle contamination.
If min./max. limit values are pre-programmed, the CS
switches the filtration unit on/off to maintain the
cleanliness within the limit value range.
Once the target cleanliness has
been reached (5x undershooting of
the ), the number of set
test cycles appears on the display (
). The test cycles (
) run. One test cycle = 60
seconds.
During this time, the last measured value will be exported via the analog
output.
After the test cycle time has elapsed, the switching output is closed and a
measurement is started.
If the result is below the reactivation threshold ( ), the inspection
cycles ( ) will start to run again. If the measured value is above this,
the switching output will remain closed until the target cleanliness (
) is no longer exceeded.