Setting the measuring mode
Mode M4: Filter to continuously monitor cleanliness class
Application:
Control of stationary offline filtration unit
Data output:
Display & RS485 & analog output & switching output
Purpose:
Establish continuous monitoring of cleanliness class
between min./max. limit values.
Function:
Control of a filter unit, continuous measurement of solid
particle contamination.
If min./max. limit values are pre-programmed, the CS
switches the filtration unit on/off to maintain the
cleanliness within the limit value range.
Once the target cleanliness has
been reached (5x undershooting of
the ), the number of set
test cycles appears on the display (
). The test cycles (
) run. One test cycle = 60
seconds.
During this time, the last measured value will be exported via the analog
output.
After the test cycle time has elapsed, the switching output is closed and a
measurement is started.
If the result is below the reactivation threshold ( ), the inspection
cycles ( ) will start to run again. If the measured value is above this,
the switching output will remain closed until the target cleanliness (
) is no longer exceeded.
T 1
ContaminationSensor CS1000
en(us)
Page 32 / 136
BeWa CS1000 3764916d 300 en-us 2017-09-25.docx 2017-09-25