SWEep Subsystem
Copyright © ITECH Electronic Co., Ltd. 97
Argument
<NRf+>[,<NRf+>,<NRf+>]
Query syntax
SWEep:VOLTage:STOP? [A|B|C][MAXimum|MINimum|DEFault]
Returns
<NR2>
Example
SWEep:VOLTage:STOP 90 //Set the stop voltage of sweep to 90V.
SWEep:VOLTage:STOP? //Query the presently set stop voltage.
SWEep:VOLTage:STOP? MAX //Query the maximum value of stop voltage.
SWEep:VOLTage:STEP <NRf+>[,<NRf+>,<NRf+>]
This command is used to set and query the step voltage of sweep.
<NRf+>: step voltage setting value, or step MINimum or MAXimum or DEFault
setting.
Note:
In single-phase (including AC, ACDC, DC, DCAC), there is only one parameter;
in three-phase, three parameters must be taken.
Syntax
SWEep:VOLTage:STEP <NRf+>[,<NRf+>,<NRf+>]
Argument
<NRf+>[,<NRf+>,<NRf+>]
*RST
0
Query syntax
SWEep:VOLTage:STEP? [A|B|C][MAXimum|MINimum|DEFault]
Returns
<NR2>
Example
SWEep:VOLTage:STEP 1 // Set the step voltage of sweep to 1V.
SWEep:VOLTage:STEP? // Query the presently set step voltage.
SWEep:VOLTage:STEP? MAX // Query the maximum value of step voltage.