IEEE-488 Common Commands
Copyright © ITECH Electronic Co., Ltd. 141
Returns
Manufacture, model, serial number, UI ver-DSP1 ver-DSP2 ver-PFC ver-Interface
ver
Example
ITECH, M7722, 00000000000004, 1.01-1.00-1.0-1.1-1.2
*OPC
This command sets the Operation Complete (OPC) bit in the Standard Event Status
Register to 1 when all other commands are complete.
Syntax
*OPC
Arguments
None
Query syntax
*OPC?
Returns
<NR1>
*RST
This command resets the power supply to default settings.
Syntax
*RST
Arguments
None
*SRE <NR1>
This command sets or queries the bits in the Status Byte Enable Register. Setting
this parameter can determine which byte of the Status Byte Register has a value
of 1. The byte sets the RQS bit of the Status Byte Register to 1. The bit definition
of the Status Byte Enable Register is as the same as the Status Byte Register.
Syntax
*SRE <NR1>
Arguments
0~255