SOURce Commands
Copyright © ITECH Electronic Co., Ltd. 35
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude][:A
C] <NRf+>[,NRf+][,NRf+]
This command is used to set and query the AC voltage RMS.
In the three-phase mode, the three parameters of the command will be set to the
corresponding A/B/C three-phase. If the command contains one Arguments, the
A/B/C will be set at the same time (in this case, whether it is balanced or
unbalanced,all so).
Syntax:
VOLTage[:LEVel][:IMMediate][:AMPLitude][:AC] <NRf+>[,NRf+][,NRf+]
Arguments:
MAXimum|MINimum|DEFault
Query syntax:
VOLTage[:LEVel][:IMMediate][:AMPLitude][:AC]? [MAXimum|MINimum|DEFault]
Returns:
MAXimum|MINimum|DEFault
Example:
VOLTage 220 //Under single mode, set the Voltage RMS to 220V
VOLTage 20,30,40 //Under multi-channel or 3-three phase mode, set the
current RMS to 30A,20A,10A
[SOURce:]VOLTage:LIMit:HIGH <NRf+> [,CH1|CH2|CH3]
This command sets and query the upper limit of RMS voltage.
Note: the second parameter is optional and only valid in multi-channel mode.
Syntax:
VOLTage:LIMit:HIGH <NRf+> [,CH1|CH2|CH3]
Arguments:
MAXimum|MINimum|DEFault
Query syntax:
VOLTage:LIMit:HIGH? [CH1|CH2|CH3],[MAXimum|MINimum|DEFault]
Returns:
MAXimum|MINimum|DEFault
Example:
VOLTage:LIMit:HIGH 300 //Set the voltage upper limit to 300V
VOLTage:LIMit:HIGH 300,1 //Set the voltage upper limit of CH1 or A phase to