SOURce Commands
Copyright © ITECH Electronic Co., Ltd. 42
Note: the second parameter is optional and valid in multi-channel
mode/three-phase.
Syntax:
POWer:LIMit:LOW <NRf+> [,CH1|CH2|CH3]
Arguments:
MAXimum|MINimum|DEFault
Query syntax:
POWer:LIMit:LOW? [CH1|CH2|CH3][,MAXimum|MINimum|DEFault]
Example
POWer:LIMit:LOW 10 //Set the lower limit of the power protection value to 10VA.
POWer:LIMit:LOW 10,1 //Set the lower power limit of channel 1 to 10VA.
POWer:LIMit:LOW? 1 //Query the lower power limit of channel 1.
[SOURce:] POWer:LIMit:DELay <NRf+> [,CH1|CH2|CH3]
This command sets and queries the power protection delay time, the unit is: s.
Syntax:
POWer:LIMit:DELay <NRf+> [,CH1|CH2|CH3]
Arguments:
MAXimum|MINimum|DEFault
Query syntax:
POWer:LIMit:DELay? [CH1|CH2|CH3][,MAXimum|MINimum|DEFault]
Example
POWer:LIMit:DELay 3 //The delay time of over power protection is set to 3s
POWer:LIMit:DELay 3,2 //Set the over-power protection delay time of
channel 2 to 3 seconds.
POWer:LIMit:DELay? 2 // Query the over-power protection delay time of
channel 2.
[SOURce:]ANGLe:STARt[:IMMediate] <NRf+>
Set and query the starting angle in normal mode.
<NRf+>: starting angle setting value
Note: the starting angle is only meaningful for phase A in the three-phase mode,
and the angle of phase B/C is set by the angle difference.
Syntax:
ANGLe:STARt[:IMMediate] <NRf+>