Auto Fill Incr
Function Group
FREQUENCY, POWER
Menu Map
2, 5
Description
This softkey is used in two lo cations:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
.
Flatness Menu - When selected, the swept signal generator waits for
a frequency incrementvalue to be entered.
--> Increment:
is
displayed in the active entry area. A list of frequencies is created
automatically, b eginning at the auto ll start frequency and always
ending with the auto ll stop frequency. The swept signal generator
uses the incrementvalue on all points, but if the stop frequency
requires a dierent increment to b e used to b e exact, the swept
signal generator simply ends the frequency list at the stop frequency
disregarding the incrementvalue.
If the incrementvalue requested creates a list that exceeds the
number of elements available, the follo wing message app ears:
TOO MANY CORRECTION PTS REQUESTED
List Menu - When selected, the swept signal generator waits for
a frequency incrementvalue to be entered.
--> Increment:
is
displayed in the active entry area. A list of frequencies is created
automatically, with all p oints separated by the frequency increment
value. The list b egins at the auto ll start frequency and ends at a
frequency less than or equal to the auto ll stop frequency
.
If the incrementvalue requested creates a list that exceeds the
number of points available (801), the following message appears:
TOO MANY LIST PTS REQUESTED
Programming Codes
SCPI:
NONE, see
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
or
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
Analyzer:
NONE
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
List Menu
\Optimizing Swept Signal Generator Performance" in Chapter 1
A-24 Operating and Programming Reference