Programming Codes
SCPI:
FREQuency:STARt
<
num
>
[freq sux] or
MAXimum
j
MINimum
j
UP
j
DOWN
FREQuency:MODE SWEep
Analyzer:
FA
<
num
>
[Hz
j
Kz
j
Mz
j
Gz]
See Also
4
CENTER
5
,
4
CW
5
, FREQUENCY
4
MENU
5
,
4
SPAN
5
,
4
STOP
5
\CW Operation and Start/Stop Frequency Sweep" in Chapter 1
\Programming Typical Measurements" in Chapter 1
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Start=M1 Stop=M2
Function Group
MARKER
Menu Map
3
Description
This softkey changes the swept signal generator start frequency
to the frequency value of marker 1, and the stop frequency to the
frequency value of marker 2. The swept signal generator has been
reset to these start/stop values, unlike the softkey
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
M1--M2 Sweep
that changes the start/stop values only while active.
Programming Codes
SCPI:
SWEep[:FREQuency]:MARKer:XFER
Analyzer:
SHMP
See Also
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
M1--M2 Sweep
\Marker Operation" in Chapter 1
S-62 Operating and Programming Reference