Leveling with Power
Meters
Leveling with a p ower meter is similar to leveling with a dio de
detector. Figure 1-13 shows the setup for power meter leveling.
Figure 1-13. Leveling with a Power Meter
1. Set up the equipment as shown. Be sure to set the p ower meter to
manual range mo de and note the range.
2. Refer to menu map 1.
3. Press
4
ALC
5
.
4.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Leveling Point PwrMtr
.
5.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pwr Mtr Range
.Enter the range value set for the power
meter as noted in step 1.
6.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Coupling Factor
, press
4
0
54
dB(m)
5
.
Unlike detector leveling, power meter leveling provides calibrated
power out of the leveled RF port.
Hint
To obtain atness corrected p ower, refer to \Creating and Applying
the User Flatness Correction Array" in the \Optimizing Swept Signal
Generator Performance" section.
Getting Started Advanced 1-27

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals