How to View or Change an GPIB Address from the Front Panel
Note
To set a GPIB address from the front panel, the instrument address
on the rear panel GPIB switch (Figure 3-2)
must
be set to 31 (all
1s).
1. Press SYSTEM
4
MENU
5
.
2.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
GPIB Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adrs Menu
.
3. The swept signal generator displays the three address softkeys:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
8360 Adrs
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meter Adrs
, and
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Printer Adrs
.
4. Select the desired softkey.
5. The swept signal generator displays the address selected for that
instrument.
6. If you wanttochange the address, use the keypad to enter the
desired address (0 to 30), then press
4
ENTER
5
.
Remember
If the swept signal generator displays
Rear panel GPIB address
must be 31 (11111) in order to change current address (=XX)
,
the address on the rear panel GPIB switch (Figure 3-2) is set to
something other than 31 (all 1s).
How to Prevent a Front Panel Change to a GPIB Address
To disable the address softkeys, set the instrument address on the
rear panel GPIB switch ( Figure 3-2) to any address
other than
31 (all
1s).
How to Set the GPIB Address on a Sw
ept Signal Generator without a
Front Panel
If your swept signal generator do es not have a front panel, set the
address on the rear panel GPIB switch (Figure 3-2) to the address
you want (factory default is 19).
Mating Connectors
All of the externally mounted connectors on the instrument are
discussed in \CONNECTORS" in Chapter 2. If you are interested
in the part numb er for a connector, see \Replaceable P
arts" in
the
Agilent Technologies 8360 B-Series Swept Signal Generator/
8360 L-Series Swept CW Generator Service Guide
.
10 MHz Frequency
Reference Selection
and Warmup Time
Tokeep the internal timebase frequency reference oven at op erating
temperature, the swept signal generator must b e connected to ac line
power. The swept signal generator requires approximately 30 minutes
to warm up from a cold start b efore the
OVEN
display message
go es o. With a stable outside temp erature, internal temp erature
equilibrium is reached after approximately two hours. For additional
information on warmup times, see Chapter 2c, \Specications".
3-8 Installation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals