SCPI COMMAND SUMMARY
Combining the Sweep Mode With the Sweep Generation Command
to Obtain the Desired Sweep Condition
Description of
Sweep Condition
:FREQ :POW :SWE :LIST :SWE
:MODE :MODE :MODE :MODE :GEN
CW Non-swept CW FIX ignored ignored ignored
Analog freq sweep SWE FIX AUTO ignored ANAL
Manual analog freq sweep SWE FIX MAN ignored ANAL
Stepp ed freq sweep SWE FIX AUTO ignored STEP
Manual step freq sweep SWE FIX MAN ignored STEP
CW with analog p ower
sweep
CW SWE AUTO ignored ANAL
CW with manual analog
power sweep
CW SWE MAN ignored ANAL
CW with stepp ed p ower
sweep
CW SWE AUTO ignored STEP
CW with manual stepp ed
power sweep
CW SWE MAN ignored STEP
Analog frequency and
power sweep
SWE SWE AUTO ignored ANAL
Manual analog frequency
and p ower sweep
SWE SWE MAN ignored ANAL
Stepp ed frequency and
power sweep
SWE SWE AUTO ignored STEP
Manual stepp ed frequency
and p ower sweep
SWE SWE MAN ignored STEP
List sweep LIST ignored ignored AUTO ignored
Manual list sweep LIST ignored ignored MAN ignored
SWEep:GENeration STEPped|ANALog
SWEep:GENeration?
Sets and queries the type of sweep to b e generated: an analog sw
eep
or a digitally stepped sweep. In either case, all of the other sw
eep
subsystem functions apply. *RST is
ANALog
.
SWEep:MANual:POINt <num>|MAXimum|MINimum
SWEep:MANual:POINt? [MAXimum|MINimum]
Sets and queries the step p ointnumber to go to and lo ck. The value
is a unitless value that is limited b etween 1 and the number of points
requested. This command has no eect on the instrument unless:
the sweep mo de is set to manual and the sweep generation is set to
stepped mo de. *RST value is 1.
SWEep:MANual[:RELative] <num>
SWEep:MANual[:RELative]?
Sets and queries a p ercentofsweep to go to and lo ck. This command
has no eect unless: the sweep mo de is set to manual and the sweep
generation is set to analog. *RST value is 0.50.