SCPI COMMAND SUMMARY
SWEep:MARKer:STATe ON|OFF|1|0
SWEep:MARKer:STATe?
Sets and queries the state of marker sweep. When
ON
, the frequency
sweep limits are taken to be the p ositions of marker 1 and marker 2.
*RST value is 0.
SWEep:MARKer:XFER
This transfers the values of marker 1 and marker 2 frequencies into
start and stop frequency.
SWEep:MODE AUTO|MANual
SWEep:MODE?
Selects and queries the manual sweep mo de switch.
AUTO The sweep is under the control of the
INIT
and
SWEEP
subsystems.
MANual FREQ:MANual, SWEep:MANual[:RELative], and
SWEep:MANual:POINt control the output.
*RST value is
AUTO
.
SWEep:POINts <num>|MAXimum|MINimum
SWEep:POINts? [MAXimum|MINimum]
Sets and queries the number of p oints in a step sweep. When points
is changed,
SWEep:STEP
is mo died by the equation
STEP = SPAN/POINTS.
Span is normally an indep endentvariable but is changed to STEP
2
POINTS if b oth of these parameters are c
hanged in the same
message. *RST value is 11.
SWEep:STEP <num>[freq suffix]|MAXimum|MINimum
SWEep:STEP? [MAXimum|MINimum]
Sets and queries the size of each frequency step.
:STEP
is governed
by the equation
STEP = SPAN/POINTS.
If you change step size then the number of points will b e changed
to span/step and a
Parameter Bumped
execution error is reported.
If span or points are changed, then STEP= SPAN/POINTS. The
step sweep command creates a coupling with sweeptime also.
If the number of p oints is changed through this coupling and
DWELl:AUTO
is
ON
and
TIME:AUTO
is
ON
, then dwell is changed to
SWEEPTIME/POINTS. Span is normally an independentvariable
but is changed to STEP
2
POINTS if b oth of these parameters are
changed in the same message. *RST value is StopMax-StartMin/10.
SWEep:TIME <num>[time suffix]|MAXimum|MINimum
SWEep:TIME? [MAXimum|MINimum]
Sets and queries the currentsweep time. The dwell time can be
coupled to sweep time if
SWE:DWEL:AUTO
is
ON
. The dwell time is
Operating and Programming Reference S-51