Pulse Rate
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Pulse Period
Function Group
4
MOD
5
Menu Map
4
Description
This softkey lets you set a value for the internal pulse generator's
pulse perio d. The range of acceptable values is from 2

s to 65.5 ms.
The factory preset value is 2 ms. When this feature is active, its
currentvalue is displayed in the active entry area.
Programming Codes
SCPI:
PULSe:TIMing:PERio d
<
num
>
[time sux] or
MAXimum
j
MINimum
Analyzer:
NONE
See Also
4
ALC
5
,
4
MOD
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse Menu
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Pulse Rate
Function Group
MODULATION
Menu Map
4
Description
This softkey lets you set the internal pulse generators pulse rep etition
rate. The rep etition rate can vary from 15.26 Hz to 500 kHz. The
factory preset value is 500 Hz. When this feature is active, its current
value is displayed in the active entry area.
Programming Codes
SCPI:
PULse:FREQuency
<
num
>
[freq sux] or
MAXimum
j
MINimum
Analyzer:
NONE
See Also
4
ALC
5
,
4
MOD
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Pulse Menu
Operating and Programming Reference P-21

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals