TRACE EXCEPTION
: This can only b e caused byaninternal
pro cessor error. Refer to Chapter 4 for instructions on contacting a
qualied service technician.
TRAP0 EXCEPTION
: This can only b e caused byaninternal
pro cessor error. Refer to Chapter 4 for instructions on contacting a
qualied service technician.
TRAP3 EXCEPTION
: This can only b e caused byaninternal
pro cessor error. Refer to Chapter 4 for instructions on contacting a
qualied service technician.
To o many test patches !!
: This error will only o ccur if the service
adjustment menu is accessed. Specically, the maximum number of
test patches has b een reached and can accept no more.
WAIT|SAVING CALIBRATION
: This error will only o ccur if the
service adjustment menu is accessed. Specically,asave calibration
has b een initiated and not yet completed when another request is
made.
WRONG PASSWORD
: This error o ccurs when the service
adjustment menu password is entered incorrectly or the wrong
password has b een used. Qualied service technicians, refer to
\ADJUSTMENTS," in the
Agilent Technologies 8360 B-Series Swept
Signal Generator/8360 L-Series Swept CW Generator Service Guide
for more information.
SCPI Error
Messages in
Numerical Order
Swept Signal Generator
Specific SCPI Error
Messages
0, No Error
: This message indicates that the device has no errors
and is currently ready to perform the op erations for whichit is
designed.
1, FUNCTION DISABLED
: The particular function invoked has
been disabled by a calibration constant. If you need access to the
function, contact a qualied service technician.
2, Wrong password
: This error o ccurs when the service adjustment
menu password is entered incorrectly or the wrong password has been
used. Qualied service technicians, refer to \ADJUSTMENTS," in
the
Agilent Technologies 8360 B-Series Swept Signal Generator/
8360 L-Series Swept CW Generator Service Guide
for more
information.
4, Unable to store data in EEROM
5, Not allowed to change address
Error Messages 2a-5