Page #428 background image
Internal Modulation
Generator
Option 002
AM, FM
Internal Waveforms:
sine, square, triangle, ramp, noise
Rate
Range
Sine: 1 Hz to 1 MHz
Square, triangle, ramp: 1 Hz to 100 kHz
Resolution: 1 Hz
Depth, deviation
Range: same as base instrument
Resolution: 0.1%
Accuracy: same as base instrument
Pulse
Mo des:
free-run, gated, triggered, delayed
Perio d range:
300 ns to 400 ms
Width Range:
25 ns to 400 ms
Resolution:
25 ns
Accuracy:
5ns
Video delay
Internal sync pulse: 0 to 400 ms
Externally-supplied sync pulse: 225 to 400 ms
Modulation Meter
Accuracy
(rates

100 kHz): 5% of range